Chapter 9 _______________________________________ Calibration, Adjustment, and Maintenance

For DMT340 M-sensor, humidity references of 0 % (for example
Nitrogen) and 10 ... 20 % are required.

NOTE

For the DMT340 M-sensor both reference humidities must be below
20 %RH.

To make the adjustment, follow the instructions below:

1.

Turn off DM70.

2.

Connect the cable 211339 between DMT340 (service port) and
DM70 (connector I or II). Turn on both the devices.

3.

With DMT340 series transmitters with the M-sensor, perform
the manual purge. Also, if you are using DMP74B probe as
reference, perform the manual purge on the reference probe as
well.

4.

Press the ADJ button on the DMT340 motherboard to enable the
adjustment mode.

5.

From the list of parameters, select RH

I/II

6.

You are prompted to check the settings, make sure that the
pressure units of DM70 and DMT340 are the same. If not,

change the unit by pressing

Unit on the corresponding row.

7.

Continue with the adjustment by pressing

Adjust. Select 2-

point adjustment. A note is displayed on the screen reminding
you to follow the adjustment procedure described in the User's

Guide. Press

OK to proceed with the adjustment.

8.

Set the probe to a lower reference value of relative humidity

1)

.

You can follow the stabilization from the graph display by

pressing

Graph. When the reading has stabilized, press

Ready.

9.

Use the arrow buttons to give the lower reference value of

relative humidity. Then press

OK.

10.

Set the probe to a higher reference value of relative humidity

1)

.

When the reading has stabilized, press

Ready. You can follow the stabilization from the graph

display by pressing

Graph.

11.

Use arrow buttons to give the higher reference value of relative

humidity. Then press

OK.

VAISALA _______________________________________________________________________ 69