Manual sensor purge, Figure 10, Sensor purge displays – Vaisala DM70 User Manual

Page 33

Advertising
background image

Chapter 5 _____________________________________________________________ User Interface

3. Select

Measurement settings and press

.

4. Select

Purge interval and press

Set.

5.

Set the desired purge interval with the arrow buttons. Press

OK to confirm.

6.

To return to the basic display, press

Exit.

Manual Sensor Purge

To check that the sensor purge has been performed to ensure the best
possible response times or before calibration, start the purge manually
as follows:

NOTE

Although the sensor purge is in the automatic mode, it can also be
performed manually.

1.

Open the menu by pressing

Open.

2. Select

Functions and press

.

3. Select

Purge and press

Start. Confirm the selection by

pressing Yes.

4.

When the sensor purge is completed, the display returns to the
basic display automatically. The stabilization of the temperature
reading can take a few minutes.

NOTE

If the dewpoint is low, automatic auto-calibration will take place
immediately after the sensor purge.

The purge will take up to 5 minutes to complete. During the purge,
there are no readings on the display. A message appears on the display
every 15 seconds to inform of the time required to complete the purge.
The purge symbol also appears in the upper left corner of the display.
See Figure 10 on page 31.

0403-044

Figure 10

Sensor Purge Displays

VAISALA _______________________________________________________________________ 31