Missing readings, Ssage is described in table 10 on – Vaisala WS425 User Manual

Page 42

Advertising
background image

USER'S GUIDE____________________________________________________________________

42 __________________________________________________________________ M210361EN-E

Table 10

The RS-232 19-character Fixed Length Output
Message

Character
position

Message

1 “

☻” 02H (<STX>, start of transmission)

2 “W”
3

“5” for 5-second running average

4

“P” for “pass”
“F” for “fail”

5

Wind direction (most significant digit)

6

Wind direction (middle digit)

7

Wind direction (least significant digit)

8

Wind speed (most significant digit)

9

Wind speed (next digit)

10

Wind speed (next digit)

11

Wind speed (least significant digit)

12

“.” (dot character)

13

Wind speed (tenth digit)

14

“M” for miles per hour; “K” for knots
“L” for kilometers per hour; “T” for meters per second

15

Check sum (most significant digit) (See note)

16

Check sum (least significant digit)

17 “

♥” 03H (<ETX>, end of transmission)

18

[CR] (carriage return)

19

[LF] (line feed)

NOTE

The checksum is calculated from 13 characters from position 2 through
14. The accumulator initializes at 0 with the addition of the byte value.
The checksum has a range of 0H … FFH.

Missing Readings

If data is missing due to a measurement problem, Handar RS-232
message reports 999.9 for wind speed.

NOTE

The sensor has a 20 ms timeout in receiving characters. Therefore,
polling strings should be transmitted by a programmable device, not as a
user command via terminal sessions.