Multi-station patient room applications – Aiphone 83873900 0602 E User Manual

Page 22

Advertising
background image

– 22 –

WIRING DIAGRAMS

Multi-Station Patient Room Applications

1. Four bed application covered by four NH-1SA/A sub stations

with a common bathroom pullcord

Multiple NH-1SA/A and NH-2SA/A bedside sub stations can also be installed
within a patient room. Observe the following conditions;
• Max. 6 bedside sub stations may be connected per room for max. 12 calls (when NH-2SA/A stations are used).
• A corridor lamp inside (NHR-4A) must not be connected with more than 4 bedside sub stations.
• Due to the above, install two NHR-4A corridor lamps when more than 4 sub stations are in one

room, i.e. two WHITE lamps used in one NHR-4A, and one RED bulb used in the other NHR-4A.

• NH-1SA/A and NH-2SA/A station numbers must be programmed in numerical order and

connected to the same subs trunk line.

• NHR-7A bathroom pullcord is optional.

NH-1SA/A

NH-1SA/A

NH-1SA/A

NH-1SA/A

1
2
3
4
5
6

WL

RL

E

TC

TL

E

ZL

LC

1
2
3
4
5
6

WL

RL

E

TC

TL

E

ZL

LC

1
2
3
4
5
6

WL

RL

E

TC

TL

E

ZL

LC

1
2
3
4
5
6

WL

RL

E

TC

TL

E

ZL

LC

Z1

11
12
13
14
15
16

3P(or 4P)

3P

2

3P(or 4P)

3 or 4

3

3P

2

3P

2

NHR-4A

NHX-80X
CCU

NHR-7A
(common
bathroom)

NHR-4A

NHR-7A

Brown
Orange
Red

Orange

Brown

Red

Orange

Yellow

Green

Blue

Purple

Brown

Orange

Purple

Gray

Brown

Orange

Purple

Gray

Brown

Orange

Purple

Gray

Yellow

Green

Blue

Blue

(*1)

(*2)

CN1

CN2

To Subs
block 1

Each pair: polarized

TO next patient room

Subs trunk line 1

The above example shows a patient room with four beds, in which a common bathroom pullcord is used.
Multiple sub applications can be arranged in the following manner.

Beds in Room

2

3

4
5
6

Subs combinations

One NH-2SA/A or two NH-1SA/A
One each NH-2SA/A & NH-1SA/A
or three NH-1SA/A
Two NH-2SA/A or four NH-1SA/A
Two NH-2SA/A & one NH-1SA/A
Three NH-2SA/A

Beds in Room

7
8
9

10
11
12

Subs combinations

Three NH-2SA/A & one NH-1SA/A
Four NH-2SA/A
Four NH-2SA/A & one NH-1SA/A
Five NH-2SA/A
Five NH-2SA/A & one NH-1SA/A
Six NH-2SA/A

(*1) Terminal Z1 for lighting Lamp 1 in NHR-3A-4.
(*2) NHR-7A may be wired to any sub station convenient. Terminal RL is wired

only on the sub station which NHR-7A is connected to.

Note; Bedside sub station Models: NH-1SA and NH-2SA CANNOT be used in multiple bed patient rooms.