13

D. Procedure to Select a Solution

NOTE: Check display to see if solution displayed (KCl, NaCl, 442 or

User) is already the type desired. If not:
1.

Press

,

or

to select the parameter on which

you wish to change the solution type.

2.

Press and hold

key

Figure 1

KCl

442

NaCl

User

for 3 seconds to make “SEL

appear (see Figure 1). For

demonstration purposes, all

4 solution types are shown

simultaneously.

3.

Use the

or

key to select type of solution desired

(ref. Solution Characteristics, pg. 40). The selected solution

type will be displayed: KCl, NaCl, 442 or User.

4.

Press

to accept new solution type.

E. Application of User Solution Type

1. User Programmable Temperature Compensation

(Tempco)

This feature allows you to change your Ultrameter II’s temperature

compensating factor to another factor between 0-9.99%/°C (ref.

Temperature Compensation, pg. 37). This feature does not apply to pH

or ORP.

a.

As in D. Procedure to Select a Solution above, select “User”

mode.

b.

With User mode now selected, press

. You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or

.

See example in Figure 2.

c.

Press

twice to skip

Figure 2

°C

% /

User

COND

calibration adjustment and

accept the new tempco (3

times if in TDS mode). You

are now ready to measure samples with your new temperature

compensation factor.