D. pH and ORP Practices to Maintain Calibration (6Pfc)

1.

Keep the sensor wet with Myron L Storage Solution.

2.

Rinse away caustic solutions immediately after use.

ORP calibration solutions are caustic, and ±5% is considered very

accurate. By using the pH zero setting (0 mV = 7 pH) for ORP and

precision electronics for detection, the Ultrameter II delivers better

accuracy without calibration than a simpler instrument could using

calibration solutions.

IX.

MEMORy

This feature allows up to 100 readings with their temperatures to be

stored simultaneously for later recall. at the same time, the TIME and

DATE are also recorded. To download the memory to a computer, (ref.

bluDock™ WIRELESS DaTa TRaNSFER INSTRUCTIONS, pg. 30).

A. Memory Storage

Figure 9

°C

KCl

COND

MEMORY

1.

While displaying a

measurement, press

to record the

displayed value.

2.

MEMORy” will appear

and the temperature

display will be momentarily replaced by a number (1-100)

showing the position of the record. Figure 9 shows a reading

of 1806 µS stored in memory record #4.

B. Memory Recall

1.

Press any measurement key.

2.

Press

, “MEMORy” will appear, and the display will

show the last record stored.

3.

Press

or

to scroll to the record location desired

(the temperature display alternates between temperature

recorded and location number).

4.

Press

to display time and date stamp.

5.

Press any measurement key to leave memory recall or allow to

automatically turn off.