Myron L 6Pfc and 4P User Manual

Page 43

Advertising
background image

39

7%

Chart 2

55

(1)%
(2)%

0%

1%

2%

3%

4%

5%

6%

0 5 10 15 20 25 30 35 40 45 50

Temperature

NaCl error with KCl tempco

442 error with KCl tempco

E. Other Solutions

A salt solution like sea water or liquid fertilizer acts like NaCl. An internal

correction for NaCl can be selected for greatest accuracy with such

solutions. Many solutions are not at all similar to KCl, NaCl or 442. a

sugar solution, or a silicate, or a calcium salt at a high or low temperature

may require a User value peculiar to the application to provide readings

close to the true compensated conductivity.

Clearly, the solution characteristics should be chosen to truly represent

the actual water under test for rated accuracy of ±1%. Many industrial

applications have historically used relative measurements seeking a

number to indicate a certain setpoint or minimum concentration or trend.

The Ultrameter II gives the user the capability to collect data in “KCl

conductivity units” to compare to older published data, in terms of NaCl

or 442, or as appropriate. The Ultrameter II can be used to reconcile

data taken with other compensation assumptions, especially with its

ability to allow custom characteristics through the User mode.

XXI. CONDUCTIVITy CONVERSION to TOTAL

DISSOLVED SOLIDS (TDS)

Electrical conductivity indicates solution concentration and ionization

of the dissolved material. Since temperature greatly affects ionization,

conductivity measurements are temperature dependent and are

normally corrected to read what they would be at 25°C (ref. Temperature

Compensation, pg. 37).