Myron L 6Pfc and 4P User Manual

Page 39

Advertising
background image

35

XVIII. TROUBLEShOOTING ChART

Symptom

Possible Cause

Corrective Action

No display, even though

measurement key pressed

Battery weak or not connected.

Check connections or replace battery. Ref. Battery Replacement, pg. 32.

Inaccurate ph readings (6Pfc)

1. pH calibration needed. Ref. pH Cal.,

pg. 17.

2. Cross-contamination from residual pH

buffers or samples in sensor well.

3. Calibration with expired pH buffers.

1. Recalibrate instrument.

2. Thoroughly rinse sensor well.

3. Recalibrate using fresh buffers. Ref. pH Buffer Solutions, pg. 36.

No response to ph changes (6Pfc)

Sensor bulb is cracked or an

electromechanical short caused by an

internal crack.

Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 37.

Will not adjust down to ph 7 (6Pfc)

pH/ORP sensor has lost KCl.

Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 32) and recalibrate. If no

improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 37).

ph readings drift or respond slowly to

changes in buffers/samples or “FAC

is displayed repeatedly (6Pfc)

1. Temporary condition due to memory” of

solution in pH sensor well for long periods.

2. Bulb dirty or dried out.

3. Reference junction clogged or coated.

Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 32) and recalibrate. If no

improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 37).

Unstable Conductivity/TDS/

Resistivity readings

1. Dirty electrodes.

2. actual resistance is changing due to

atmospheric contamination.

1. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 32.

2. Minimize test sample exposure to air by taking a flowing sample.

Ref. Measuring Resistivity, pg. 9.

Unable to calibrate Conductivity/

TDS

Film or deposits on electrodes.

Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 32.

Resistivity readings much lower than

expected

1. Contamination from previous sample or

from pH sensor well.

2. Carbon dioxide in test sample.

1. Rinse cell cup more thoroughly before measurement. Ensure pH cap is snugly in

place.

2. See Measuring Resistivity, pg. 9.

Low ORP Reading

Slow or no response to ORP changes

(6Pfc)

ORP platinum electrode is dirty.

Check the ORP sensor functioning. Take an ORP reading of Myron L pH/ORP

Sensor Storage Solution (ref. pH Sensor Storage Solution (6Pfc), pg. 36). If the

reading is outside the range of 350-400 mV, clean ONLY the platinum ORP

electrode with Myron L ORP Conditioner solution-soaked cotton swab (ref. ORP

Sensor Conditioner Solution (6Pfc), pg. 36), being careful not to touch the swab to

the glass bulb of the pH sensor.

Advertising