Myron L 6Pfc and 4P User Manual

Page 24

Advertising
background image

20

VIII.

CALIBRATION INTERVALS

There is no simple answer as to how often one should calibrate an

instrument. The Ultrameter II is designed to not require frequent

recalibration. The most common sources of error were eliminated in

the design, and there are no mechanical adjustments. Still, to ensure

specified accuracy, any instrument must be checked against chemical

standards occasionally.

A. Suggested Intervals

On the average, we expect calibration need only be checked monthly

for the Conductivity, RES or TDS functions. The pH (6Pfc) function

should be checked every 2 weeks to ensure accuracy. Measuring some

solutions will require more frequent intervals.

B. Calibration Tracking Records

To minimize your calibration effort, keep records. If adjustments you

are making are minimal for your application, you can check less often.

Changes in conductivity calibration should be recorded in percent.

Changes in pH calibration (6Pfc) are best recorded in pH units.
Calibration is purposely limited in the Ultrameter II to ±10% for the

conductivity cell, as any change beyond that indicates damage, not

drift. Likewise, calibration changes are limited to ±1 pH unit (6Pfc), as

any change beyond that indicates the end of the sensor’s lifetime and

replacement is recommended.

C. Conductivity, RES, TDS Practices to Maintain Calibration

1.

Clean oily films or organic material from the cell electrodes

with foaming cleaner or mild acid. Do not scrub inside the cell.

2.

Calibrate with solutions close to the measurements you make.

Readings are compensated for temperature based on the type

of solution. If you choose to measure tap water with a KCl

compensation, which is often done (ref. an Example of 2

different solution selections and the resulting compensation,

pg. 38), and you calibrate with 442 solution because it is

handy, the further away from 25°C you are, the more error you

have. Your records of calibration changes will reflect

temperature changes more than the instrument’s accuracy.

3.

Rinse out the cell with pure water after taking measurements.

Allowing slow dissolving crystals to form in the cell

contaminates future samples.

4.

For maximum accuracy, keep the pH sensor cap on tight so

that no fluid washes into the conductivity cell.

Advertising