Sets the probe the parameter is measured from.

2

Sets the parameter that is measured.

3

Sets the Warning levels for the parameter.* A parameter measured outside the
warning levels is stored in the log memory even if the Tube is not designated to
save that reading. Level values should use the same units as used by the probes.

4

Sets the Alarm levels for the parameter. A parameter measured outside the Alarm
levels is stored in the log memory even if the Tube is not designated to save that
reading, and an alarm is sent as long as the alarm has been enabled and satisfies
the exceedences rules.* Level values should use the same units as used by the
probes.

5

Sets the alarm mode.
Absolute mode triggers warnings and alarms when the measured value is lower
than the minimum value or higher than the maximum value.
Incremental mode triggers warnings and alarms when the difference between two
consecutive measurements is lower than the minimum value or higher than the
maximum value.

6

Writes settings to the Tube.

*For more information on warnings and alarms, see "Alarms - Overview" on page 40.