Setting up a log, 3 configuring the telemetry device, Overview – In-Situ Tube 300R Telemetry System User Manual

Page 9: Opening the unit, Configuring the telemetry device

Advertising
background image

800-446-7488

9

www.in-situ.com

Instrument Communication Settings - TROLL 9500 Only

1. Connect to the instrument using Win-Situ 4 software.

2. Right click on the TROLL 9500 device in the tree on the left and select Edit.

3. Select ASCII Mode Preferences and click Next.

4. Select the following options:

Comma delimited
Enable ASCII mode of linear tests

5. Click Finish.

For the TROLL 9500 to function, you must set up a log and set the logging interval to be
the same as the Tube logging interval.

Setting Up a Log

The Tube logs data from instrument directly to the internal SD card. How and when the
data is recorded is configured using the Tube set up tool.

If you require data redundancy, program a log on your instrument prior to connecting it
to the Tube. Some instrument do not support logs.

Refer to your instrument manual for instructions on setting up a log.

Configuring the Telemetry Device

Overview

If you purchased a data plan through In-Situ Inc., you need to open the unit and connect
the battery, and then set up the Tube using the configuration software.

If you are using a 3rd party data plan, you will need to install the SIM card prior to
setting up the Tube with the configuration software. See "SIM Card" on page 16.

Opening the Unit

1. Place the unit on a flat surface.

2. Remove the metal ring from the top of the unit .

3. Locate the small round pin housed between the black o-ring and the outer body

threads. Insert a flathead screwdriver next to the pin and rotate the screwdriver until
the internal fixture comes free. Remove the internal fixture from the outer body.