Absolute mode – In-Situ Tube 300R Telemetry System User Manual

Page 41

Advertising
background image

800-446-7488

41

www.in-situ.com

When a Warning is triggered, the measurement is stored in the log memory even if the
Tube is not designated to save that reading. When an Alarm is triggered, the reading is
stored and an alarm message is sent (if the corresponding alarm has been enabled).

Alarm Mode is set in the Probe and Set-Point Configuration Menu, under the
Parameters Monitoring tab.

Absolute Mode

In Absolute Mode, alarms and warnings occur when a measurement value is less than
the minimum threshold value or greater than the maximum threshold value.

Minimum Level Value

Maximum Level

Value

Measurement Value

Result

1

15

16

ALARM

1

15

1

OK

1

15

15

OK

1

15

0

ALARM

-2

10

1

OK