Calibrate:data, Calibrate:level, Calibrate:password – Agilent Technologies 6631B User Manual

Page 45: Calibrate:save

Advertising
background image

Language Dictionary - 4

45

CALibrate:DATA

This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.

Command Syntax

CALibrate:DATA<NRf>

Parameters

<external reading>

Unit

A (amperes)

Examples

CAL:DATA 3222.3 MA CAL:DATA 5.000

Related Commands

CAL:STAT CAL:LEV

CALibrate:LEVel

This command selects the next point in the calibration sequence.

P1: the first calibration point
P2: the second calibration point

Command Syntax

CALibrate:LEVel <point>

Parameters

P1 | P2

Examples

CAL:LEV P2

CALibrate:PASSword

This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE.

If the password is set to 0, password protection is removed and the ability to enter the calibration mode is
unrestricted.

Command Syntax

CALibrate:PASScode<NRf>

Parameters

<model number> (default)

Examples

CAL:PASS 6812 CAL:PASS 6.1994

Related Commands

CAL:SAV

CALibrate:SAVE

This command saves any new calibration constants after a calibration procedure has been completed in
nonvolatile memory. If CALibrate:STATe OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored..

Command Syntax

CALibrate:SAVE

Parameters

None

Examples

CAL:SAVE

Related Commands

CAL:PASS CAL:STAT