MEASure:CURRent?
FETCh:CURRent?

FETCh:CURRent? applies to Agilent 66312A, 66332A Only

These queries return the dc output current.

Query Syntax

MEASure:[SCALar]:CURRent[:DC]?
FETCh:[SCALar]:CURRent[:DC]?

Parameters

None

Examples

MEAS:CURR? MEAS:CURR:DC?

Returned Parameters

<NR3>

Related Commands

MEAS:VOLT?

MEASure:CURRent:ACDC?
FETCh:CURRent:ACDC?

Agilent 66312A, 66332A Only

These queries return the ac+dc rms output current.

Query Syntax

MEASure:[SCALar]:CURRent:ACDC?
FETCh:[SCALar]:CURRent:ACDC?

Parameters

None

Examples

MEAS:CURR:ACDC? FETC:CURR:ACDC?

Returned Parameters

<NR3>

Related Commands

MEAS:VOLT:ACDC?

MEASure:CURRent:HIGH?
FETCh:CURRent:HIGH?

Agilent 66312A, 66332A Only

These queries return the High level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 1024 bins between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data points in the high bin is
returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points,
then the maximum value is returned by these queries.

Query Syntax

MEASure[SCALar]:CURRent:HIGH?
FETCh[:SCALar]:CURRent:HIGH?

Parameters

None

Examples

MEAS:CURR:HIGH?FETC:CURR:HIGH?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR:LOW? CALC:REF:HIGH