Compatibility language - b 87 – Agilent Technologies 6631B User Manual

Page 87

Advertising
background image

Compatibility Language - B

87

STATUS

1

2

16

64

128

CV

+CC

UNR

OT

not used

OC

ERR

INH

OV

4

512

-CC

FAST

8

32

256

1024

2048

REGISTER

NORM

STATUS

1

2

16

64

128

4

512

8

32

256

1024

2048

REGISTER

ACCUMULATED

MASKUS

1

2

16

64

128

4

512

8

32

256

1024

2048

REGISTER

CV

+CC

UNR

OT

not used

OC

ERR

INH

OV

-CC

FAST

NORM

FAULT

1

2

16

64

128

4

512

8

32

256

1024

2048

REGISTER

SERIAL

1

2

16

64

128

4

8

32

REGISTER

POLL

not used

not used

not used

PON

FAU

RDY

ERR

RQS

Figure B-1. COMpatibility Status Model

Table B-4. Bit Assignment of Status, Astatus, Fault, & Mask Registers

Bit Position

11

10

9

8

7

6

5

4

3

2

1

0

Bit Name

NORM

FAST

-CC

INH

ERR

OC

not
used

OT

OV

UNR

+CC

CV

Bit Weight

2048

1024

512

256

128

64

32

16

8

4

2

1

CV = The unit is operating in constant voltage mode.
CC+ = The unit is operating in constant current mode.
UNR = The output of the unit is unregulated.
OV = The overvoltage protection circuit has tripped.
OT = The over-temperature protection circuit has tripped.
OC = The overcurrent protection circuit has tripped.
ERR = A programming error has occurred. Use ERR? to clear.
CC = The unit is operating in negative constant current mode.
INH = The external remote inhibit signal has turned the output off.
FAST = The output is in Fast operating mode. (Agilent 66332A, 6631B, 6632B, 6633B, 6634B only)
NORM = The output is in Normal operating mode. (Agilent 66332A, 6631B, 6632B, 6633B, 6634B only)

Table B-5. Bit Configuration of Serial Poll Register

Bit Position

7

6

5

4

3

2

1

0

Bit Name

not used

RQS

ERR

RDY

not used

not used

PON

FAU

Bit Weight

64

32

16

2

1

RQS = The dc source has generated a service request. Use a serial poll to clear.
ERR = Same as ERR bit in Status register. Use ERR? to clear.
RDY = This bit cleared when unit busy processing commands. Set when processing complete.
PON = A Power-on has occurred. Use CLR to clear.
FAU = A bit has been set in the Fault register. Use FAULT? to clear.