SYSTem:LOCal

For RS-232 Operation Only

This command places the dc source in local mode during RS-232 operation. The front panel keys are
functional.

Command Syntax

SYSTem:LOCal

Parameters

None

Example

SYST:LOC

Related Commands

SYST:REM SYST:RWL

SYSTem:REMote

For RS-232 Operation Only

This command places the dc source in remote mode during RS-232 operation. This disables all front
panel keys except the Local key. Pressing the Local key while in the remote state returns the front panel to
the local state.

Command Syntax

SYSTem:REMote

Parameters

None

Example

SYST:REM

Related Commands

SYST:LOC SYST:RWL

SYSTem:RWLock

For RS-232 Operation Only

This command places the dc source in remote mode during RS-232 operation. All front panel keys
including the Local key are disabled. Use SYSTem:LOCal to return the front panel to the local state.

Command Syntax

SYSTem:RWLock

Parameters

None

Example

SYST:RWL

Related Commands

SYST:REM SYST:LOC

*IDN?

This query requests the dc source to identify itself. It returns a string composed of four fields separated by
commas.

Query Syntax

*IDN?

Returned Parameters

<AARD>

Field

Information

Agilent Technologies

Manufacturer

xxxxxA

model number followed by a letter suffix

nnnnA-nnnnn

10-character serial number or 0

<A>.xx.xx

Revision levels of firmware.

Example

AGILENT,66312A,0,A.00.01