Language Dictionary - 4

77

TRIGger:SEQuence2:LEVel:CURRent
TRIGger:ACQuire:LEVel:CURRent

Agilent 66312A, 66332A Only

This command sets the trigger level for internally triggered current measurements. A positive current
trigger occurs when the current level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative current trigger occurs when the
current level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:LEVel:CURRent<NRf+>
TRIGger:ACQuire:LEVel:CURRent<NRf+>

Parameters

0 to MAX (see table 4-3)

Unit

A (amperes)

*RST Value

0

Examples

TRIG:SEQ2:LEV:CURR 5 TRIG:ACQ:LEV:CURR MAX
TRIG:ACQ:LEV 2

Query Syntax

TRIGger:SEQuence2:LEVel:CURRent?
TRIGger:ACQuire:LEVel:CURRent?

Returned Parameters

<NR3>

Related Commands

TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:HYST:CURR

TRIGger:SEQuence2:LEVel:VOLTage
TRIGger:ACQuire:LEVel:VOLTage

Agilent 66312A, 66332A Only

This command sets the trigger level for internally triggered voltage measurements. A positive voltage
trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the
voltage level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.

Command Syntax

TRIGger:SEQuence2:LEVel:VOLTage<NRf+>
TRIGger:ACQuire:LEVel:VOLTage<NRf+>

Parameters

0 to MAX (see table 4-3)

Unit

V (volts)

*RST Value

0

Examples

TRIG:SEQ2:LEV:VOLT 5 TRIG:ACQ:LEV:VOLT MAX
TRIG:ACQ:LEV 2

Query Syntax

TRIGger:SEQuence2:LEVel:VOLTage?
TRIGger:ACQuire:LEVel:VOLTage?

Returned Parameters

<NR3>

Related Commands

TRIG:SEQ2:LEV:CURR TRIG:SEQ2:HYST:VOLT