Trigger, Trigger:source, Trigger:sequence2 trigger:acquire – Agilent Technologies 6631B User Manual

Page 74

Advertising
background image

4 - Language Dictionary

74

TRIGger

When the transient trigger subsystem is initiated, this command generates a trigger signal. The trigger will
then:

1. Initiate a pending level change as specified by CURRent:TRIGger or VOLTage;TRIGger.
2. Clear the WTG bit in the Status Operation Condition register after both transient and acquire trigger

sequences have completed. (WTG is the logical-or of both transient and acquire sequences.)

3. If INITiate:CONTinuous ON has been programmed, the trigger subsystem is immediately re-enabled

for subsequent triggers. As soon as it is cleared, the WTG bit is again set to 1.

Command Syntax

TRIGger[:SEQuence1][:IMMediate]
TRIGger[:TRANsient][:IMMediate]

Parameters

None

Examples

TRIG TRIG:IMM

Related Commands

ABOR CURR:TRIG INIT *TRG VOLT:TRIG

TRIGger:SOURce

This command is included for completeness. It selects the trigger source for transient triggers. Since BUS
is the only trigger source for transient triggers, this command does not need to be used.

BUS

GPIB device, *TRG, or <GET> (Group Execute Trigger)

Command Syntax

TRIGger[:SEQuence1]:SOURce<source>
TRIGger[:TRANsient]:SOURce<source>

Parameters

BUS

*RST Value

BUS

Examples

TRIG:SOUR BUS

Query Syntax

TRIGger[:SEQuence1]:SOURce?
TRIGger[:TRANsient]:SOURce?

Returned Parameters

<CRD>

TRIGger:SEQuence2
TRIGger:ACQuire

Agilent 66312A, 66332A Only

When the trigger subsystem is initiated, these commands generate a measurement trigger signal. The
measurement trigger causes the dc source to measure the output voltage and current and store the
results in a buffer.

.

Command Syntax

TRIGger:SEQuence2[:IMMediate]
TRIGger:ACQuire:[:IMMediate]

Parameters

None

Examples

TRIG:SEQ2 TRIG:ACQ

Related Commands

TRIG:SOUR TRIG:SEQ2:DEF TRIG:SEQ2:COUN
TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:SLOP:CURR