Measure:current:low? fetch:current:low, Measure:current:maximum? fetch:current: maximum, Measure:current:minimum? fetch:current:minimum – Agilent Technologies 6631B User Manual

Page 49

Advertising
background image

Language Dictionary - 4

49

MEASure:CURRent:LOW?
FETCh:CURRent:LOW?

Agilent 66312A, 66332A Only

These queries return the Low level current of a current pulse waveform. The instrument first measures the
minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 1024 bins between the maximum and minimum data points. The bin containing the most
data points below the 50% point is the low bin. The average of all the data points in the low bin is returned
as the Low level. If no low bin contains more than 1.25% of the total number of acquired points, then the
minimum value is returned by these queries.

Query Syntax

MEASure[SCALar]:CURRent:LOW?
FETCh[:SCALar]:CURRent:LOW?

Parameters

None

Examples

MEAS:CURR:LOW? FETC:CURR:LOW?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR:HIGH? CALC:REF:LOW

MEASure:CURRent:MAXimum?
FETCh:CURRent: MAXimum?

Agilent 66312A, 66332A Only

These queries return the maximum output current reading from the measurement sample.

Query Syntax

MEASure[:SCALar]:CURRent:MAXimum?
FETCh[:SCALar]:CURRent:MAXimum?

Parameters

None

Examples

MEAS:CURR:MAX? FETC:CURR:MAX?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR:MIN?

MEASure:CURRent:MINimum?
FETCh:CURRent:MINimum?

Agilent 66312A, 66332A Only

These queries return the minimum output current reading from the measurement sample.

Query Syntax

MEASure[:SCALar]:CURRent:MINimum?
FETCh[:SCALar]:CURRent:MINimum?

Parameters

None

Examples

MEAS:CURR:MIN? FETC:CURR:MIN?

Returned Parameters

<NR3>

Related Commands

MEAS:CURR:MAX?

Advertising
This manual is related to the following products: