4 - Language Dictionary

44

Calibration Commands

Calibration commands let you:

u

Enable and disable the calibration mode

u

Change the calibration password

u

Calibrate the current and voltage programming and measurement, and store new calibration
constants in nonvolatile memory.

NOTE:

If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error.

CALibrate:CURRent

This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.

Command Syntax

CALibrate:CURRent[:SOURce][:DC][:POSitive]

Parameters

None

Examples

CAL:CURRCAL:CURR:SOUR:DC:POS

Related Commands

CAL:CURR:NEG

CALibrate:CURRent:NEGative

This command initiates the calibration of the negative dc output current.

Command Syntax

CALibrate:CURRent[:SOURce][:DC]:NEGative

Parameters

None

Examples

CAL:CURR:NEGCAL:CURR:SOUR:DC:NEG

Related Commands

CAL:CURR

CALibrate:CURRent:MEASure:LOWRange

This command initiates the calibration of the low-range current measurement circuit.

Command Syntax

CALibrate:CURRent:MEASure[:DC]:LOWRange

Parameters

None

Examples

CAL:CURR:MEAS

Related Commands

CAL:CURR

CALibrate:CURRent:MEASure:AC

Agilent 66312A, 66332A Only

This command initiates the calibration of the high bandwidth (ac) measurement circuit.

Command Syntax

CALibrate:CURRent:MEASure:AC

Parameters

None

Examples

CAL:CURR:MEAS:AC