6 Operation

BioPhotometer Data Transfer Software — Operating Manual

3. Confirm your entry with OK.

The input window is closed. The assigned identification code for the following measurement
is shown In ID in the information field of the Work area.

ID-allocation remains active for all successive measurements until one of the following conditions
applies:

ID is changed manually with the icon.

The method is changed.

The user is changed (only with active user administration).

The software is exited.

During a method change the following query appears. An ID can be used for different methods.

6.6.2

Gather data simultaneously for measurement at the BioPhotometer plus.

Requirement
The desired method was not yet called at the BioPhotometer plus.



Select at the BioPhotometer plus the corresponding method and start with your measurement
at BioPhotometer plus.

The measured data is automatically transferred and displayed in the Work area. There data can
be displayed in DataOutput or as a table under Samples. (see Display files on page 16).

6.6.3

Transfer data from the results memory to the BioPhotometer plus

1. Select the corresponding measurement result in the BioPhotometer plus results memory.
2. Press Enter on the BioPhotometer plus in order to transfer the measurement result.

The data is transferred and the date is marked with an * in the area Samples and Calibrations
(as well as in the Work area and in the Archive area).

Hint!

For the correct transfer of all information from measurement series you must call the method at
BioPhotometer plus before transferring data of the first measurement result.If alternatively the
data transfer is started during measurement series it leads to a faulty display of the date.