MX-5 DL Ultrasonic Thickness Gauge

11

If the Stability Indicator has fewer than five bars darkened, or the

numbers on the display seem erratic, first check to make sure that there is

an adequate film of couplant beneath the transducer, and that the

transducer is seated flat against the material. If the condition persists, it

may be necessary to select a different transducer (size or fre quency) for

the material being measured. See page 30 for information on transducer

selection.

While the transducer is in contact with the material that is being

measured, the MX-5 DL will perform four measurements every second,

updating its display as it does so. When the transducer is removed from

the surface, the display will hold the last measurement made.

IMPORTANT

Occasionally, a small film of couplant will be drawn out between the

transducer and the surface as the transducer is removed. When this

happens, the MX-5 DL may perform a measurement through this couplant

film, resulting in a measurement that is larger or smaller than it should be.

This phenomenon is obvious when one thickness value is observed while

the transducer is in place, and another value is observed after the

transducer is removed.