Dakota Ultrasonics MX-5 DL User Manual

Page 19

Advertising
background image

MX-5 DL Ultrasonic Thickness Gauge

15

Calibration to a known thickness

NOTE: This procedure requires a sample piece of the specific

material to be measured, the exact thickness of which is known,

e.g. from having been measured by some other means.

1) Make sure the M X-5 DL is on.

2) Perform a Probe-Zero (refer to page 13)

3) Apply couplant to the sample piece.

4) Press the transducer against the sample piece, making sure that

the transducer sits f lat against the surface of the sample. The

display should show some (probably incorrect) thickness value,

and the Stability Indicator should have nearly all its bars on.

5) Having achieved a stable reading, remove the transducer. If the

displayed thickness changes from the value shown while the

transducer was coupled, repeat step 4.

6) Press the CAL key. The IN (or MM) symbol should begin flashing.

7) Use the UP and DOWN arrow keys to adjust the displayed

thickness up or down, until it matches the thickness of the sample

piece.

8) Press the CAL key again. The IN/

µs (or M/s) symbols should

begin flashing. The MX-5 DL is now displaying the sound velocity

value it has calculated based on the thickness value that was

entered in step 7.

9) Press the CAL key once more to exit the calibration mode. The

MX-5 DL is now ready to perform measurements.