20

In normal operation, the MX-5 DL performs and displays four

measurements every second, which is quite adequate for single

measurements. In Scan Mode, however, the tool performs sixteen

measurements every second. While the transducer is in contact with the

material being measured, the MX-5 DL is keeping track of the lowest

measurement it finds. The transducer may be "scrubbed" across a surface,

and any brief interruptions in the signal will be ignored. When the

transducer loses contact with the surface for more than a second, the MX-5

DL will display the smallest measurement it found.

When the MX-5 DL is in measurement mode, press the MODE key to

activate the features and settings. The display will begin flashing ALAr

(alarm), followed by the status - on/off. Use the UP and DOWN arrow keys

to scroll to SCAn mode. Press the SEND key to toggle scan mode on/off.

While scanning, the display is updated faster than normal which may result

in flickering of the values displayed. When the transducer is removed from

the material being scanned, the MX-5 DL will (after a brief pause) display

the smallest measurement it found.

Alarm Mode

The Alarm Mode feature of the MX-5 DL allows the user to set an

audible and visual parameter when taking measurements. If the

measurement falls below a nominal value, set by the user, a red light will

be illuminated on the front panel of the gauge and the beeper sounded.

This improves the speed and efficiency of the inspection process by

eliminating constant viewing of the actual reading displayed. The following

procedures outline how to enable and set up this feature: