MX-5 DL Ultrasonic Thickness Gauge

23

7) Press the SEND key to select the desired nominal value and return

to mode menu.

8) Press the MODE key once again to return to measurement mode.

RS232 Serial Port & The Data Logger

The MX-5 DL is equipped with an RS232 serial port. Using the

accessory cable (part# N-306-0010), the MX-5 DL has the ability to connect

to a computer, or external storage device. The following section outlines

the procedure for connecting the MX-5 DL to a computer, and how to

collect data using any standard communications program:

Connecting To a Computer

1) Connect the accessory cable (part# N-306-0010) to the 2 pin jack

located on the bottom of the MX-5 DL, and the 9 pin connector to a

serial port on the computer.

2) Start the communications software that will be used to collect the

measurements (i.e. Microsoft Windows

3.1 -Terminal, or 95 -

HyperTerminal).

3) Setup the communications software using the following

parameters: Data Bits - 8, Parity - None, Stop Bits - 1, Baud

Rate 1200 (to print a report), or 9600 to transfer data file.

§ Note: A report can be printed to a communications program (i.e.

HyperTerminal), or printed to a serial printer using 8.5" x 11" paper.