Lower Workspace Load Cell Connection

8. Install the appropriate fixturing for your test specimen.

9. Set the crosshead limit stops.

a) Ensure that the crosshead is stationary and that the test parameters are set.
b) Set the crosshead limit stops.

For tension testing, set the stop slightly above the expected maximum upward crosshead travel.

For compression testing, set the stop slightly above the test starting point.

c) Tighten the stop securely on the limit rod.
d) Set the lower limit stop.

For tension testing, set the stop just below the starting position.

For compression testing, set the stop just below the expected maximum downward crosshead
travel.

e) Tighten the stop securely on the limit rod.

122 | MTS Criterion

Series 40

Best Practices for Other System Configurations and Tests