Resetting and silencing alarms – RKI Instruments GX-2003 Manual User Manual

Page 30

Advertising
background image

23 • Measuring Mode, Alarms

GX-2003 Operator’s Manual

Resetting and Silencing Alarms

You can set the GX-2003’s gas alarms as latching or self-resetting
alarms (see “Updating the Alarm Latching Setting” on page 56).

• Self-resetting alarms (ALARM LATCHING set to OFF)

Self-resetting alarms automatically shut off and reset when the gas
reading falls below (or rises above for an oxygen low alarm) the
alarm setting. You cannot silence or reset self-resetting alarms.

• Latching alarms (ALARM LATCHING set to ON)

You can set latching alarms with or without Alarm Silence (see
“Updating the Alarm Silence Setting” on page 56).

With ALARM SILENCE On and ALARM LATCHING On:

When the GX-2003 goes into gas alarm, press the RESET SILENCE
button to silence the buzzer. If the gas concentration was still above

Over Range*

• Gas reading indicates

full scale.

• Alarm LED arrays

flash.

• Back light turns on.

• Pulsing tone once per

second

• Unit vibrates once per

second

Low Flow*

• The display indicates

FAIL LOW FLOW
LEVEL

• The Alarm LED’s flash

• Rapid pulsing tone

• Unit vibrates every

few seconds

Low Battery
Warning*

• Battery icon blinks

• None

Dead Battery Alarm*

• Gas readings

replaced by Change
the Batteries

• Alarm LED arrays

flash

Rapid pulsing tone
alternating with vibration

Sensor Failure*

FAIL SENSOR

appears at the top of
the display and the
failed sensor(s) are
indicated

• Alarm LED arrays

flash

Rapid pulsing tone
alternating with vibration

*This alarm can also occur in Bar Hole Mode and Leak Check Mode

Table 4: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications