Updating the confirmation beep setting, Updating the interval time setting, Updating the datalog overwrite setting – RKI Instruments GX-2003 Manual User Manual

Page 64

Advertising
background image

57 • Using Setup Mode

GX-2003 Operator’s Manual

Updating the Confirmation Beep Setting

With CONFIRMATION BEEP set to ON, the GX-2003 beeps once
every 15 minutes to verify that it is operating.

With CONFIRMATION BEEP set to OFF (factory setting), the GX-
2003 does not sound a confirmation beep.

1. From the main menu, place the cursor next to the

CONFIRMATION BEEP

menu item. Press and release the

POWER ENTER button.

2. Use the ▲AIR or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Interval Time Setting

This setting indicates how often the GX-2003 records readings into the
data logger. The following interval times can be selected: 10 minutes; 5
minutes; 3 minutes; 1 minute; 30 seconds; 20 seconds; 10 seconds. The
factory setting is 5 minutes.

1. From the main menu, place the cursor next to the INTERVAL

TIME

menu item. Press and release the POWER ENTER button.

2. Use the ▲AIR or (SHIFT) ▼ buttons to display the desired setting.

3. Press and release the POWER ENTER button to SAVE the setting

and return to the main menu.

Updating the Datalog Overwrite Setting

With DATALOG OVERWRITE set to ON (factory setting), the GX-
2003 writes over the oldest data with new data when the data logger
memory is full.

With DATALOG OVERWRITE set to OFF, the GX-2003 stops saving
data to the data logger when the data logger memory is full.

CONFIRMATION BEEP

OFF

21:09

INTERVAL TIME

5 MIN

21:09

Advertising