GX-2003 Operator’s Manual

Overview • 68

Chapter 7: Maintenance

Overview

This chapter describes troubleshooting procedures for the GX-2003. It
also includes procedures for recharging the battery pack and
replacing various consumable parts.

WARNING: RKI Instruments, Inc. recommends that service,

calibration, and repair of RKI instruments be performed
by personnel properly trained for this work. Replacing
sensors and other parts with original equipment does
not affect the intrinsic safety of the instrument.

Troubleshooting

The troubleshooting table describes error messages, symptoms,
probable causes, and recommended action for problems you may
encounter with the GX-2003.

Table 6: Troubleshooting the GX-2003

Symptoms

Probable

Causes

Recommended Action

• The LCD is blank.

• The unit may

have been turned
off.

• The alkaline

batteries may
need to be
replaced or the
Ni-MH battery
pack recharged.

1. To turn on the unit, press and

briefly hold the POWER
ENTER button.

2. If the unit does not turn on,

replace the alkaline batteries
or recharge the Ni-MH battery
pack.

3. If the difficulties continue,

contact RKI Instruments, Inc.
for further instruction.

• Display indicates

“Circuit Trouble”
during startup.

• A circuit fault has

occurred on one
of the PCB’s.

• Contact RKI Instruments, Inc.

for further instruction.

• The LCD shows

abnormally high or
low readings but
other gas
detection
instruments do
not.

• The GX-2003

may need to be
recalibrated.

• The sensor for

the affected
channel(s) may
need
replacement.

1. Recalibrate the unit.
2. If the difficulties continue,

replace the sensor for the
affected channel(s) and
calibrate the affected
channel(s).

3. If the difficulties continue,

contact RKI Instruments for
further instruction.