Updating the lcd back light time setting, Turning the password function on or off – RKI Instruments GX-2003 Manual User Manual

Page 72

Advertising
background image

65 • Using Setup Mode

GX-2003 Operator’s Manual

Updating the LCD Back Light Time Setting

This setting indicates the length of time the LCD illuminates when
you press any button. The minimum setting is 0 seconds; the
maximum setting is 60 seconds. The factory setting is 30 seconds.

1. From the main menu, place the cursor next to the LCD BACK

LIGHT TIME

menu item. Press and release the POWER ENTER

button.

2. Use the ▲AIR and (SHIFT)▼ buttons to display the desired

setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Turning the Password Function On or Off

With PASSWORD ON/OFF set to ON, the GX-2003 prompts you for a
password when you enter Calibration Mode or Setup Mode.

With PASSWORD ON/OFF set to OFF (factory setting), no password
is required to enter Calibration Mode or Setup Mode.

1. From the main menu, place the cursor in front of the PASSWORD

ON/OFF

menu item. Press and release the POWER ENTER button.

2. Use the ▲AIR or (SHIFT)▼ button to display the desired setting.

3. If you selected OFF, press and release the POWER ENTER button

to save the setting and return to the main menu.
If you selected ON, press and release the POWER ENTER button.
ON

is replaced by four characters with the first character being a

flashing 0 and the rest asterisks (*).

4. Use the ▲AIR or (SHIFT)▼ button to select a number from 0 to 9.

5. Press and release the POWER ENTER button to enter the selection

and advance to the next character.

LCD BACK LIGHT

TIME

30 SEC

21:09

SET

PASSWORD

OFF

21:09

Advertising