Display Mode • 26

• display time in operation

• display TWA and STEL readings (H

2

S and CO only)

• display the date, time, and battery charge

• clear the data logger

• display remaining log time

To enter Display Mode, press and release the DISPLAY (ADJ) button
while in Measuring Mode. To scroll from one screen to the next press
and release the DISPLAY (ADJ) button.

NOTE:

Each screen displays for 20 seconds. If you do not press the
DISPLAY (ADJ) button to scroll to the next screen within 20
seconds, the GX-2003 automatically returns to Measuring
Mode.

HC Range Screen

This screen displays only if your GX-2003 is equipped with both the catalytic
% LEL combustible sensor and the TC % volume combustible sensor.
It
allows you to select the display units for the combustible channel as
% LEL/% volume autoranging or % volume. If autoranging is
selected, the CH4 reading will be displayed in % LEL up to 100% LEL,
which is equivalent to 5% volume for methane (CH

4

). Above 100%

LEL, it will be displayed in %volume. If % volume is selected, the CH

4

reading will be displayed in % volume only.

CAUTION:

If the combustible channel is set for % volume only, then there
are no gas alarms for any detection channel.

To update the HC range setting:

1. After entering Display Mode and arriving at the HC range screen,

press and release the ▲AIR button. The current setting will flash at
the bottom of the LCD.

2. Use the ▲AIR or (SHIFT)▼button to change the setting.

3. Press the POWER ENTER button to save the change and advance

to the next screen.

HC Range

YES : AIR

NO : DISPLAY