GX-2003 Operator’s Manual

Using Setup Mode • 58

1. From the main menu, place the cursor next to the DATALOG

OVERWRITE

menu item. Press and release the POWER ENTER

button.

2. Use the ▲AIR or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER/ENTER button to save the setting

and return to the main menu.

Updating the Date and Time Settings

1. From the main menu, place the cursor next to the DATE/TIME

menu item. Press and release the POWER ENTER button. The date
and time will be displayed with the year flashing.

2. Use the ▲AIR or (SHIFT)▼ button to display the desired year.

3. Press and release the POWER ENTER button to save the setting.

The month setting flashes.

4. Repeat steps 2 and 3 to enter the month, day, hours, and minutes

settings. The main menu displays after you enter the minutes
setting.

Updating the Calibration Interval

This setting defines the amount of time between calibrations. The
minimum setting is 1 day and the maximum setting is 255 days. The
factory setting is 90 days.

1. From the main menu, place the cursor next to the CAL.

INTERVAL

menu item. Press and release the POWER ENTER

button.

DATA LOGGER

OVERWRITE

ON

21:09

SET DATE/TIME

Apr/01/’03

21:09

21:09

SET CALIBRATION

INTERVAL

90 DAY

21:09