2.4 Device connection at the MSB

14

Titrando Installation Instructions

plifiers can be connected in series; this allows a maximum extension of

25 m.

Start the PC Control or tiamo software.

The Titrando will be recognized automatically. When the PC Control or

tiamo software is started a system test will be carried out automatically

on the Titrando. The LED "On" on the Titrando lights up when the sys-

tem test is finished and the instrument is ready for use.

2.4 Device connection at the MSB

The following devices can be connected via each of the four MSB con-

nections (Metrohm Serial Bus): one stirrer or titration stand, one Do-

simat or Dosino dosing device and one remote box. The stirrers and

remote boxes each have an MSB output so that the devices can be

switched sequentially ("daisy chain"). The following illustration provides

an overview of the devices that can be connected to an MSB.

Fig. 8: Overview of MSB connections

The MSB 1 of the Titrando with internal dosing drive is occupied by the

internal dosing drive. This means that only a stirrer and a remote box

can be connected to MSB 1.
First connect all peripheral devices and then connect the Titrando to the

mains supply. MSB connections can be extended with the 6.2151.010

cable. The maximum length of the connection is 15 m.
The Titrando automatically recognizes which device has been con-

nected to which MSB connection. The Touch Control or the PC software

(PC Control or tiamo) shows the connected devices in the device man-

ager or the configuration dialog respectively. All devices connected to

the MSB are operated by the Touch Control or PC Control/tiamo.