4 appendix, 1 technical data, 1 titration and measuring modes – Metrohm 835 Titrando User Manual

Page 41: 2 measuring interfaces, Appendix, Technical data, Titration and measuring modes, Measuring interfaces

Advertising
background image

4

Appendix

Titrando Installation Instructions

35

4 Appendix

In this section you will find the most important technical data of the Ti-

trando, a list of standard and optional accessories and the warranty

and conformity declarations.

4.1 Technical

data

Provided that nothing to the contrary is mentioned, the published val-

ues are typical technical data for the Titrandos with internal and external

dosing drive.

4.1.1 Titration and measuring modes

DET

Dynamic equivalence point titration

Controlled titrant addition with variable volume increments

MET

Monotonic equivalence point titration

Titrant addition with fixed volume increments

SET

Titration to one or two predefined endpoints

KFT

Karl Fischer titration

STAT

Endpoint titration while holding the measured value constant

MEAS

Measurement (pH, potential, temperature, current (Upol), poten-

tial(Ipol) and concentration)

4.1.2 Measuring

interfaces

1 or 2 galvanically separated measuring interfaces
Potentiometry
1 high-impedance measuring input for pH, redox and ISE electrodes
1 reference input for separate reference electrode
Input resistance

Offset current

> 1 * 10

12

Ohm

< 1 * 10

–12

A

(under reference conditions)

Temperature
1 measuring input for temperature sensors (Pt1000 or NTC)

Automatic temperature compensation, for NTC sensors R (25 °C) and B

25/50

can be

configured.
Polarizer
1 measuring input for polarized electrodes

Polarization current Ipol

Polarization potential Upol

–125.0 ... +125.0 µA in 2.5 µA steps

–1250 ... +1250 mV in 25 mV steps

Advertising