2.6 Sensor connection

30

Titrando Installation Instructions

2.6.2 Differential

potentiometry

During potentiometric measurements in poorly conducting media, high-

impedance measuring sensors are affected by electrostatic and elec-

tromagnetic interferences. Use our Solvotrode 6.0229.100 or any other

special electrode for pH measurements in organic solvents. If this does

not help then you can connect a Differential amplifier 6.5104.030

(230 V) or 6.5104.040 (115 V).
The Differential amplifier is connected to the high-impedance measur-

ing input (Ind.).

2.6.3 Titration

vessel

setup

In titration it is important that the solution is thoroughly mixed. The stir-

ring rate should be high enough for a small vortex to be formed. If the

stirring rate is too high then air bubbles will be entrained. This results in

incorrect measurements. If the stirring rate is too low then the solution

at the electrode will not be correctly mixed. To ensure that the meas-

urement is made in a thoroughly mixed solution after titrant addition the

buret tip should be located where the turbulence is greatest. In addi-

tion, the distance between the spot where the titrant is added and the

electrode should be as large as possible. When positioning the elec-

trode and buret tip you must also take the stirring direction into ac-

count.

1

2
3

Fig. 20: Recommended arrangement of magnetic stirring bar (1),

electrode (2) and buret tip (3)