Metrohm 835 Titrando User Manual

Page 23

Advertising
background image

2

Installation

Titrando Installation Instructions

17

"Status" LED

Dosing device operating status

off

There is no exchange unit at-

tached.

constant illumination

The Titrando is ready for dosing

or titrating. The exchange unit

has been attached correctly and

recognized and is now in the

change position, i. e. the ex-

change unit can be removed.

blinks slowly

The Titrando is currently dosing

or filling or the exchange unit is

not in the change position.

An intelligent 806 Exchange unit

has been attached and the data

on the built-in data chip is cur-

rently being read out or written.

blinks rapidly

Dosing drive error (see Sec-

tion 3, Troubleshooting)