2

Installation

Titrando Installation Instructions

19

Fig. 12: Example for connecting a dosing device:

Titrando – 805 Dosimat

The 685 Dosimat is connected to the Titrando with the 6.2134.030 ca-

ble. 685 Dosimat and 700 Dosino types must be connected directly to

the MSB socket of the Titrando (see Fig. 8: Overview of MSB connec-

tions). If you want to connect a stirrer (see Section 2.4.1) and an exter-

nal 800 Dosino (or 805 Dosimat) dosing device to a Titrando without in-

ternal dosing drive then we recommend that the stirrer is connected to

MSB 1 and the dosing device to the MSB connection of the stirrer as

this corresponds to the standard settings in the methods.
Type 806 Exchange units for the 805 Dosimat and type 807 Dosing

units for the 800 Dosino have a built-in data chip that allows the storage

of data about the exchange or dosing unit and the reagent. This data is

automatically read out or updated by Touch Control or PC Con-

trol/tiamo when the exchange or dosing unit is attached to the dosing

device. The data is edited in the Touch Control or in the PC Con-

trol/tiamo software.