4.1 Technical data

36

Titrando Installation Instructions

Polarizer (857 only)
1 measuring input for polarized electrodes

Polarization current Ipol

1)

Polarization potential Upol

2)

–122.5 ... +122.5 µA in 2.5 µA steps

–1225 ... +1225 mV in 25 mV steps

1)

–125.0 / +125.0 µA: values not guaranteed, dependent on reference potential

+2.5 V

2)

–1250 / +1250 mV: values not guaranteed, dependent on reference potential

+2.5 V

4.1.3 Specification of the measuring inputs

Measuring range

Resolution

Measuring accuracy

1)

pH

–20.000 ... +20.000

0.001 pH

± pH 0.003

Potential

2)

–2000 mV ... +2000 mV

0.1 mV

± 0.2 mV

Current

3)

–200 µA ... +200 µA

0.01 µA

Temperature

Pt1000

NTC

–150 °C ... +250 °C

–20 °C ... +250 °C

4)

0.1 °C

0.1 °C

± 0.2 °C (Pt1000: –20 °C ...

+150 °C)

1)

± 1 digit, without sensor error, under reference conditions

2)

potentiometric and voltametric

3)

amperometric

4)

for a NTC-Sensor with R (25 °C) = 30'000 Ohm and B (25/50) = 4100 K.

Measuring cycle: 100 ms for all measuring ranges

4.1.4 Specification of the measuring inputs (857 only)

Measuring range

Resolution

Measuring accuracy

1)

pH

–13.000 ... +20.000

0.001 pH

± pH 0.003

Potential

2)

–1200 mV ... +1200 mV

0.1 mV

± 0.2 mV

Current

3)

–120 µA ... +120 µA

0.01 µA

Temperature

Pt1000

NTC

–150 °C ... +250 °C

–5 °C ... +250 °C

4)

0.1 °C

0.1 °C

± 0.2 °C (–20 °C... +150 °C)

± 0.6 °C (+10 °C... +40 °C)

1)

± 1 digit, without sensor error, under reference conditions

2)

potentiometric and voltametric

3)

amperometric

4)

for a NTC-Sensor with R (25 °C) = 30'000 Ohm and B (25/50) = 4100 K.

Measuring cycle: 100 ms for all measuring ranges