Metrohm 835 Titrando User Manual

Page 40

Advertising
background image

3.1 Problems

34

Titrando Installation Instructions

Problem

Possible cause

Measures

The dosing drive is over-

loaded because the flat cock

is blocked.

Switch off the Touch Control or end

PC Control/tiamo. Check whether

the exchange unit can be removed.

If it cannot be removed then check

whether the flat cock can still be

rotated. Move it manually to the

change position by turning it to the

right (see Instructions for Use of

exchange unit). Remove the ex-

change unit and proceed as de-

scribed in the Instructions for Use

in the section "Servicing a blocked

flat cock".

The dosing drive is over-

loaded because the piston is

blocked. The fault is indi-

cated by the software (Touch

Control or PC Control/tiamo).

Switch the control device off and

then on again. The dosing device is

initialized when switching on. Re-

move the exchange unit and clean

it as described in the Instructions

for Use in the section "Care and

maintenance". Contact Metrohm

Service if it is not possible to re-

move the exchange unit.

Titrando with internal

dosing drive only:

"Status" LED blinks

rapidly.

The exchange unit data can

no longer be read as the

data chip has been dam-

aged mechanically or by

chemicals.

Have the data chip replaced by

Metrohm Service personnel.

In order to be able to use the ex-

change unit until the data chip is

replaced you can remove the data

chip yourself. The cylinder volume

will still be recognized automati-

cally, but data can no longer be

read from or stored in the ex-

change unit.