1 connecting stirrers and titration stands, Connecting stirrers and titration stands, Fig. 9: titrando – stirrer – Metrohm 835 Titrando User Manual

Page 21

Advertising
background image

2

Installation

Titrando Installation Instructions

15

Attention!

If you are operating the Titrando with the Touch Control then make

sure that the Touch Control is switched off while you are setting up or

breaking connections between the instruments. If you are operating

the Titrando with the PC software then pull out the plug from the

power supply socket before setting up or breaking MSB connections.

2.4.1 Connecting stirrers and titration stands

You can use the 801 Magnetic Stirrer, the 803 Titration Stand (stir-

ring from below) or the 804 Titration Stand with 802 Rod Stirrer (stir-

ring from above).

Connect the appropriate stirrer as follows:

Fig. 9: Titrando – Stirrer

You can connect a maximum of one 801 Stirrer, one 804 Titration Stand

with 802 Rod Stirrer or one 803 Titration Stand to each MSB socket. The

801 Stirrer as well as the 804 and 803 Titration Stands has a MSB con-

nection to which an additional device, e. g. a dosing device, can be

connected. We recommend that the stirrer is connected to MSB 1, as

this corresponds to the default setting in the methods.
The stand and the stand support are included with the stirrer or titration

stand. Fasten the stand support to the base of the Titrando with the four

screws supplied. Decide whether you want to attach the stirrer to the

right or left of the Titrando. The assembly of the support rod and the

stirrer or titration stand is described in the Instructions for Use of the

801 Stirrer and the 804 Titration Stand or the 803 Titration Stand.