Metrohm 850 Professional IC AnCat Nonsuppressed User Manual

Page 22

Advertising
background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

850 Professional IC–AnCat – non-suppressed

Place the detectors in the instrument and connect them (see
Chapter 3.6.2, page 22)
.

Remove transport locking screws (see "Removing transport lock-
ing screws", page 24)
.

Connect the leak sensor (see "Connecting the leak sensor", page
24)
.

Connect drainage tubings (see "Installing drainage tubing", page
27)
.

3 Installing the eluent path

The following steps need to be carried out once for each of the two
channels.

Equip the eluent aspiration tubing and connect with the eluent
bottle (see Chapter 3.8.1, page 30).

At the end of the pre-installed column inlet capillaries (3-2) and ,
respectively (3-5), connect one 6.2744.040 coupling each, using a
6.2744.014 pressure screw, instead of the columns.

Connect detector inlet capillaries (3-3) and, respectively (3-6), to
the 6.2744.014 coupling, using a 6.2744.040 pressure screw.

4 Installing the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

The following steps need to be carried out once for each of the two
channels.

Connect the 6.1803.040 sample aspirating capillary (3-8) connec-
ted at the sample input of the injection valve to the output of the
sample degasser with the aid of a 6.2744.090 long pressure
screw (see Chapter 3.13, page 43).

Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through.

In the cation channel, replace the 6.1825.210 sample loop with
the 6.1825.220 sample loop (see "", page 45).

Advertising