4 setting up the instrument, 1 packaging, 2 checks – Metrohm 850 Professional IC AnCat Nonsuppressed User Manual

Page 26: 3 location, Setting up the instrument, Packaging, Checks, Location, 3) and, respectively (3-6) , to, 8) connec

Advertising
background image

3.4 Setting up the instrument

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

850 Professional IC–AnCat – non-suppressed

thermostat.

3

Detector inlet capillary

4

Eluent aspiration tubing 6.1834.080
Connected to the eluent degasser.

5

Column inlet capillary
Connected to the injection valve and threa-
ded into the capillary recesses of the column
thermostat.

6

Detector inlet capillary

7

Sample aspirating capillary 6.1803.040
Connected to the injection valve. The other
end can be connected either to the sample
degasser or guided directly out of the instru-
ment.

8

PTFE connection capillary
Section of the 6.1803.040 PTFE capillary.
Connects the sample degasser with the
Sample Processor (required only if the sam-
ple degasser is used)

9

Sample aspirating capillary 6.1803.040
Connected to the injection valve. Can
optionally be connected to the sample
degasser.

10 PTFE connection capillary

Section of the 6.1803.040 PTFE capillary.
Connects the sample degasser with the
Sample Processor (required only if the sam-
ple degasser is used)

11 PEEK pressure screws, short

6.2744.070
For connecting the capillaries to the guard
or separation column, respectively.

12 Separation columns

13 Conductivity detectors

You will find detailed descriptions of the individual installation steps in the
following chapters.

3.4

Setting up the instrument

3.4.1

Packaging

The instrument is supplied in highly protective special packaging together
with the separately packed accessories. Keep this packaging, as only this
ensures safe transportation of the instrument.

3.4.2

Checks

Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.

3.4.3

Location

The instrument has been developed for operation indoors and may not be
used in explosive environments.

Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.

Advertising