Figure 28, Conductivity detector rear – Metrohm 850 Professional IC AnCat Nonsuppressed User Manual

Page 61

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC–AnCat – non-suppressed

■■■■■■■■

51

1

3

2

Figure 28

Conductivity detector rear

1

Detector cable
With mounted plug.

2

Detector output capillary
Permanently installed.

3

Type plate
With serial number.

Note

In order to prevent unnecessary peak widening after separation, the
connection between the outlet of the separation column and the inlet
in the detector should be kept as short as possible.

Connect the detector input capillary to the separation col-
umn

1 Connecting detector input

Fasten the detector input capillary (29-1) on the output of the col-
umn (29-3) using a 6.2744.070 pressure screw (29-2).