3 Installation

850 Professional IC–AnCat – non-suppressed

■■■■■■■■

13

5 Connecting the instrument

Connect the instrument to the PC (see Chapter 3.17.1, page
52)
.

Connect instrument to mains supply (see Chapter 3.17.2, page
52)
.

6 Initial start-up

(see Chapter 4.1, page 57).

Switch on the PC and start MagIC Net.

Switch on and prepare the instrument.

Deaerate the high pressure pump.

Rinse the instrument without column(s).

7 Installing the columns

The following steps need to be carried out once for each of the two
channels:

Remove the 6.2744.040 coupling between column inlet capillaries
and detector inlet capillaries.

Install the guard column (if used) (see "Connecting and rinsing
the guard column", page 54)
.

Install the separation column (see "Connecting and rinsing the
separation column", page 56)
.

Connect the detector inlet capillary (see "Connect the detector
input capillary to the separation column", page 51)
.

8 Conditioning the instrument

(see Chapter 4.2, page 58).

The instrument is now ready for measuring samples.