6 Troubleshooting
850 Professional IC–AnCat – non-suppressed
■■■■■■■■
85
Problem
Cause
Remedy
Vacuum is not being
built
Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.
■
Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:
■
Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.