2.2 Mains connection

752 Pump Unit

9

10

10

12

12

15

15

230

11

11

115

Fig. 3: Setting the mains voltage

2.2.2

Fuses

Two fuses 0.2 A/slow-blow for 100

120 V or 0.1 A/slow-blow for

220

240 V are installed in the fuse holder 10

10 of the 752 Pump Unit as

standard.

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

For checking or changing fuses, proceed as described in section 2.2.1.

2.2.3

Mains cable and mains connection

Mains cable

The instrument is supplied with one of three mains cables:

6.2122.020 with plug SEV 12 (Switzerland,

)

6.2122.040 with plug CEE(7), VII (Germany,

)

6.2133.070 with plug NEMA 5-15 (USA,

)

which are three-cored and fitted with a plug with an earthing pin. If a

different plug has to be fitted, the yellow/green lead (IEC standard)

must be connected to protective earth (protection class 1).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Mains connection

Plug the mains cable into mains connection plug 15

15 of the 752 Pump

Unit (see Fig. 3).

2.2.4

Switching the instrument on/off

The 752 Pump Unit is switched on and off using mains switch 12

12. When

the instrument is switched on lamp 33 lights up.

100 – 120 V

220 – 240 V

10

10 Fuse holder

11

11 Voltage selection

insert with display

of voltage

12

12 Mains switch

15

15 Mains connection

plug