Metrohm 752 Pump Unit User Manual

Page 15

Advertising
background image

2 Installation

752 Pump Unit

12

16

16

17

17

18

18 19

19

44

55

19

19 17

17

90

90//93

93

77

20

20

20

20

16

16 20

20

22

22 23

23 20

20

21

21

Fig. 6: Installing pump tubings

44

Tubing cartridge

20

20 PEEK compression fitting

(6.2744.010)

55

Contact pressure lever

21

21

Connector (6.2824.110) with filter

77

Snap-action lever

22

22

Housing for 6.2824.100 Filter unit

PEEK

16

16

PTFE tubing (6.1803.020)

23

23

Connector

17

17

Coupling (6.2744.030)

90

90

Suppressor inlet capillary for

H

2

SO

4

18

18

Pump tubing (6.1826.050)

93

93

Suppressor inlet capillary for H

2

O

19

19

Stopper (white-yellow)

Press contact pressure lever 55 on both tubing cartridges

down as far as it will go.

Insert a length of pump tubing 18

18 (6.1826.050) into each of

the tubing cartridges as shown in Fig. 6. The white-yellow

stopper 19

19 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

Place the tubing cartridges on mounting pin 99 and press

down on the right-hand side until snap-action lever 77 clicks

into position on holding clamp 66. Take care that no kinks are

formed in the pump tubing.

8 Suppressor connection 2: H

2

SO

4

Loosen rotary nipple screwed onto the interior side of con-

nection 27

27. Pull inlet capillary 90

90 marked with "H

2

SO

4

" (see

Fig. 17 of 732/733 Manual) by hand out of the opening of

connection 27

27 as far as required. Retighten nipple on the in-

terior side of connection 27

27 to fix inlet capillary 90

90.