10 Startup of 752 Pump Unit

Set switch <REMOTE> 22 on 752 Pump Unit to "OFF".

Switch on 752 Pump Unit with mains switch 12

12 and set switch

<PUMP> 33 to "ON".

Adjust the contact pressure for both tubing cartridges: press

contact pressure lever 55 upwards until the solutions just start

to be drawn in. Then press the contact pressure lever up-

wards until it clicks once more to obtain optimal contact pres-

sure.

Check all capillaries between the 752 Pump Unit, the sup-

pressor module and the waste container for escaping liquid. If

liquid escapes anywhere the corresponding compression fit-

ting must be tightened further or changed.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever 77 on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).