3 Operation

752 Pump Unit

16

Switch on remote control

ON

OFF

REMOTE

In order to allow remote operation via 732 IC Detector the remote

control must be switched on.

ON

Remote control switched on

OFF

Remote control switched off

Automatic pump switch-on when the 732 IC

Detector is switched on

In order to start up the drive of the 752 Pump Unit automatically

when the 732 IC Detector is switched on alterations must be

made to the basic settings (Setup, see section 4.4.1 of 732/733

Manual). The remote output lead 1 must be set to 1 (on, active,

0 V):

>SETUP/output
remote 10000000

When the 732 IC Detector is switched off the pump is also

switched off.

Switch pump on/off in a program

In the 732 IC Detector the <PROGRAM> key can be used to

create time programs with a maximum of 20 program steps (see

section 4.7.1 of 732/733 Manual). The pump can be switched on

or off at any program step by setting the remote output lead 1 to

1 (on, active, 0 V) or 0 (off, inactive, open) respectively:

Switch on:

>PROGRAM/edit XXX.X min
remote

1

TTTTTTT

Switch off:

>PROGRAM/edit XXX.X min
remote

0

TTTTTTT

Switch pump on/off at an event

In the 732 IC Detector the <EVENT> key can be used to

program a maximum of 4 different events (see section 4.7.3 of

732/733 Manual). The pump can be switched on or off at any

event by setting the remote output lead 1 to 1 (on, active, 0 V) or

0 (off, inactive, open) respectively:

Switch on:

EVENT YY-MM-DD HH:MM:SS
remote

1

TTTTTTT

Switch off:

EVENT YY-MM-DD HH:MM:SS
remote

0

TTTTTTT