3 air bubbles in pump circuit, 4 exchanging the pump tubing – Metrohm 752 Pump Unit User Manual

Page 21

Advertising
background image

4 Troubleshooting

752 Pump Unit

18

4.3

Air bubbles in pump circuit

If air bubbles can be seen in the pump circuit then the aspirating tubing

must be immersed completely in the liquid. If this has no effect then the

connections must be checked (see section 4.2).

4.4

Exchanging the pump tubing

Pump tubings are consumable material with a limited lifetime and

should be exchanged at regular intervals (approx. every 2 weeks under

continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in section 2.3.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges should be

raised completely by loosening snap-action lever 77 on the right-hand

side (the set contact pressure remains unchanged).
As the pump is always operated on the same side the pump tubings

6.1826.050 supplied can be used on both sides. To exchange a pump

tubing proceed as follows:

1 Remove old pump tubing

Press contact pressure lever 55 on the tubing cartridge down

as far as it will go.

Release tubing cartridge 44 from holding clamp 66 by pressing

down snap-action lever 77 and remove from mounting pin 99

on the 752 Pump Unit (see Fig. 1).

Remove old pump tubing.

2 Insert new pump tubing

Insert the new pump tubing 18

18 (6.1826.050) in the tubing

cartridge as shown in Fig. 6. The white-yellow stopper 19

19

must click into the corresponding holder on the left-hand side

of the tubing cartridge.

Place the tubing cartridge on mounting pin 99 and press down

on the right-hand side until snap-action lever 77 clicks into po-

sition on holding clamp 66. Take care that no kinks are formed

in the pump tubing.

3 Set contact pressure

Press contact pressure lever 55 upwards until the solution just

starts to be drawn in. Then press contact pressure lever up-

wards until it clicks once more to obtain optimal contact pres-

sure.